Syringe Pump model: KL-602

Specification

MODEL KL-602
Syringe Size
10, 20, 30, 50/60 ml
Applicable Syringe
Compatible with syringe of any standard
VTBI
0.1-9999 ml
<1000 ml in 0.1 ml increments  /  ≥1000 ml in 1 ml increments
Flow Rate
Syringe 10 ml: 0.1-400 ml/h
Syringe 20 ml: 0.1-600 ml/h
Syringe 30 ml: 0.1-900 ml/h
Syringe 50/60 ml: 0.1-1300 ml/h
<100 ml/h in 0.1 ml/h increments  /  ≥100 ml/h in 1 ml/h increments
Bolus Rate
Syringe 10 ml: 400 ml/h
Syringe 20 ml: 600 ml/h
Syringe 30 ml: 900 ml/h
Syringe 50/60 ml: 1300 ml/h
Anti-Bolus
Automatic
Accuracy ±2% (mechanical accuracy ≤1%)
Infusion Mode
Flow rate: ml/min, ml/h
Time-based
Body weight: mg/kg/min, mg/kg/h, ug/kg/min, ug/kg/h, etc.
KVO Rate
0.1-1 ml/h (in 0.1 ml/h increments)
Alarms
Occlusion, near empty, end program, low battery, end battery,
AC power off, motor malfunction, system malfunction, reminder alarm,
pressure sensor error, syringe installation error, syringe drop off
Additional Features
Real-time infused volume, automatic power switching, automatic syringe identification, mute button, purge, bolus, anti-bolus, system memory, key locker
Drug Library
Available
Occlusion Sensitivity
High, medium, low
Docking Station Stackable up to 4-in-1 or 6-in-1 Docking Station
Power Supply, AC
110/230 V (optional), 50/60 Hz, 20 VA
Battery 9.6±1.6 V, rechargeable
Battery Life
7 hours at 5 ml/h
Working Temperature
5-40℃
Relative Humidity
20-90%
Atmospheric Pressure 860-1060 hpa
Size 314*167*140 mm
Weight 2.5 kg

 

ใบรับแจ้งรายการละเอียด เลขที่ 65-2-2-2-0015877

*อ่านคำเตือนในฉลากและเอกสารกำกับเครื่องมือแพทย์ก่อนใช้*

-----------------------------------------------------------------

Warning:

  • Checking and confirming the normal operation of the pump periodically. Check the residual liquid and the syringe conditions. Don’t just rely on the pump alarm system.
  • Avoid using the pump in presence of flammable gases and flammable anesthetic mixture with air, oxygen or nitrous oxide.
  • The use of any mobile phone very close to the pump is not allowed since the high frequency noise during the conversation could cause malfunction of the pump.
  • The use of the pump in MRI rooms such as high-pressure rooms or places where high electromagnetic radiation is generated is not allowed.
  • During injection, the range of pump position shall not exceed ±65cm (based on the position of patients’ heart)

 

Precautions:

  • Please use fingers to operate this pump. Any sharp tool may damage the panel of pump.
  • If pump not used for a period of time (over one month), or the power of battery is very low, please recharge the battery by connecting to AC power supply. If the effective time reduce obviously after being charged, it show battery reach the end of life. Please contact authorized dealer to replace it.
  • In case the pump does not work as the user’s manual and can’t find the reason. Please stop the operation and report the details of the problem
  • In order to ensure the connection of injection line is stable, it is better to use mechanical syringe or spiral syringe

 

Contraindication: N/A