Electrosurgical model: Surtron 300HP/400HP
✦ TWIN OPERATIVE ACTIVITY
- Two output channels for monopolar cutting
- Two independent setting and output channels for the most frequent monopolar coagulation
✦ CONTROLLED CUTTING PROCEDURES
- Automatic adaptive system of power* (TACS - Tissue Adaptive Contact System)
- Enhanced cutting current for eschar free incision
- Cutting current with selectable coagulating property for bleeding controlled dissection
- Low capacitive coupling by slow BLEND current for endoscopic procedures
✦ TWIN CONTROLLED COAGULATIVE PROCEDURES
- SPEEDY** coagulation and DEEP***coagulation available to two operators
- SPRAY coagulation available by hand switch or foot switch controls
✦ WIDE BIPOLAR FEATURES
- Bipolar cutting current with selectable bleeding control
- Slow BLEND cutting current
- Separately activable Automatic Start/Stop coagulation
- Vessel Sealing up to 7mm
✦ SURGEON AND PATIENT TOTAL SAFETY
- Acceptability of contact impedance between patient body and return electrode continuously watched by Skin Plate Electronic Control (SPEC)
- H.F. Leakage currents continuously monitored
- Mains fluctuation automatically compensated before to start the delivery of HF power
✦ PERSONALIZED PROCEDURES
- Independent choise of coagulative currents and power levels by each of two operators
- Ten personal configurable and immediately recallable procedures
✦ EXTENTION DRIVER AND INTERLOCK SIGNALS
- Drived output connection to external Argon Plasma unit and Smoke Evacuator unit
✦ AUTOMATIC SELF DIAGNOSIS
- Automatic diagnostic control of the electronic unit and the connected accessories with coded report of failures
-------------------------------------
*Monopolar CUT mode until 100 Watt.
**SPEEDY coagulation also said Fulgurate or Forced coagulation
***DEEP coagulation also said Pin Point, Dessicate or Soft coagulation
Specification
MODEL | Surtron 300HP | Surtron 400HP |
Reference code | 10400.802 |
10400.902 |
Maximum output power monopolar pure CUT | 300 W - 300 Ω |
400 W - 300 Ω |
Maximum output power monopolar cut-coag CUT80% | 250 W - 300 Ω |
300 W - 300 Ω |
Maximum output power monopolar cut-coag CUT60% | 200 W - 300 Ω |
250 W - 300 Ω |
Maximum output power monopolar cut-coag CUT40% | 150 W - 300 Ω |
200 W - 300 Ω |
Maximum output power monopolar cut ENHANCED | 200 W - 500 Ω |
250 W - 500 Ω |
Maximum output power monopolar cut-coag BLEND | 200 W - 300 Ω |
250 W - 300 Ω |
Maximum output power monopolar coag SPEEDY | 100 W - 500 Ω |
120 W - 500 Ω |
Maximum output power monopolar coag DEEP | 100 W - 200 Ω |
120 W - 200 Ω |
Maximum output power monopolar coag SPRAY | 70 W - 2000 Ω |
100 W - 2000 Ω |
Maximum output power monopolar coag SPRAY ARGON | 70 W - 2000 Ω |
100 W - 2000 Ω |
Maximum output power bipolar CUT | 120 W - 50 Ω |
120 W - 50 Ω |
Maximum output power bipolar cut-coag CUT80% | 120 W - 50 Ω |
120 W - 50 Ω |
Maximum output power bipolar cut-coag CUT60% | 120 W - 50 Ω |
120 W - 50 Ω |
Maximum output power bipolar cut-coag CUT40% |
100 W - 50 Ω | 100 W - 50 Ω |
Maximum output power bipolar cut-coag BLEND | 120 W - 150 Ω | 120 W - 150 Ω |
Maximum output power bipolar coag COAG | 120 W - 50 Ω | 120 W - 50 Ω |
Maximum output power bipolar coag VESSEL SEALING | 200 W - 50 Ω | 200 W - 50 Ω |
Working frequency Monopolar / Bipolar | 425 kHz / 525 kHz | 425 kHz / 525 kHz |
Patient circuit | F | F |
Selectable input voltage | 115 - 230 Vac | 115 - 230 Vac |
Mains frequency | 50 - 60 Hz | 50 - 60 Hz |
Electrical input power | 1000 VA | 1000 VA |
Size WxHxD mm | 470x150x400 | 470x150x400 |
Weight | 17.5 Kgs | 17.5 Kgs |
ใบรับแจ้งรายการละเอียด เลขที่ 66-2-2-2-0010692
*อ่านคำเตือนในฉลากและเอกสารกำกับเครื่องมือแพทย์ก่อนใช้*
-----------------------------------------------------------------
Warning:
- Electro-surgery can be dangerous. Careless use of any element in the electrosurgical system may subject the patient to a serious burn. Read and understand all warnings, precautions, and directions for use before attempt touse any active electrode. Neither LED SpA, can be considered responsible for personal, material or consequential injury, loss or damage that results from improper use of the equipment and accessories.
- before switch on the unit, operator has to verify that requested mains voltage corresponds to the voltage available from the electrical net. (see chapter INSTALLATION).
Precautions: Avoid the skin to skin contact (for example between arm and body of the patient).
Contraindication:
Electro surgery is not recommended in the following subjects:
- having pacemaker;
- with stimulating electrodes;
- with metal prosthesis plant;
- with important arterial pressure unbalance;
- with important nervous disorders;
- with renal insufficiency;
- in state of pregnancy.
Burns are the most consequences of the HF electro surgery for the patient, even if these are not the only one. In fact necrosis by compression, allergic reactions to the disinfectant, gas or inflammable liquids ignition. Some important causes of burns are by:
- insufficient medical equipe training about all modalities to avoid or reduce the risks of burns by using HF electrosurgical units;
- use of disinfectants with high alcol content;
- incorrect position of the patient during the electrosurgical operation;
- contact between active electrode and the skin;
- contact with liquid;
- long application of HF currents;
- incorrect application of the patient-plate.
To avoid or reduce the common HF electrosurgical risks it is very important to respect the rules and safety measurements exposed illustrate on the next chapter.